Patient Warming Pitfalls to Avoid

One of surgery's most basic patient safety practices is riskier than you might think.

Jared Bilski

BIO

BAD IDEA
Pamela Bevelhymer, RN, BSN, CNOR
BAD IDEA Disconnecting the hose on forced-air warming units and using it to directly warm the patient or the bed can lead to serious burns and contaminate the sterile field.

Active patient warming prevents hypothermia and many related complications such as cardiac trouble, impaired immune function, increased risk of infection and lengthier post-op stays. But you have to worry about an entirely new set of safety issues when staff don't follow proper warming protocols. Unfortunately, that occurs far more often than it should.

"Warming devices aren't benign pieces of equipment designed solely for patient comfort," says Michelle Feil, MSN, RN, CPPS, WCC, clinical practice leader at Penn Medicine in Philadelphia, Pa. "These are therapeutic medical devices that carry a real risk for patient harm."

Ms. Feil would know. She authored a Pennsylvania Patient Safety Authority report that analyzed 278 harmful or potentially harmful events associated with patient warming devices (osmag.net/7yCKNk). Here are a few of the real-world warming mishaps she discovered.

  • Thermal injuries. The most frequently reported type of patient harm in Ms. Feil's report come in a variety of forms, but they nearly all stem from the same root cause: Using warming devices in a manner that's not consistent with the manufacturers' Instructions for Use (IFU). For example, the report cited an example of a patient who'd been lying directly on top of a warming blanket during an appendectomy, which resulted in reddening of his skin and a small, reddened area on the right buttock. That usage went against the blanket's IFU, which specifically stated patients should never be positioned on top of the blanket, that they should only be covered by it.

Then there's the extremely dangerous practice of "free hosing," where the hose of a forced-air warming device is placed underneath surgical drapes or the blanket of a patient's bed or stretcher. This off-label use has resulted in a number of cases of serious patient burns because of the change in the distribution of the heat.

New to Outpatient Surgery Magazine?
Sign-up to continue reading this article.
Register Now
Have an account? Please log in:
Email Address:
  Remember my login on this computer

advertiser banner

Other Articles That May Interest You

Almost Left Behind

Lessons learned from a final count that came up 2 sponges short.

Safety: Need Help Establishing a Culture of Safety?

This survey self-assessment tool will identify areas of needed improvement.

On Point: Have You Established a Culture of Safety?

Talk is cheap when it comes to protecting patients and staff from harm.