All 16 Reprocessed Ureteroscopes Tested in Study Were Still Contaminated

Researchers stress the importance of consistently monitoring reprocessing outcomes.

Published: June 15, 2017

URETEROSCOPES A study found flexible ureteroscopes were still contaminated after they'd been reprocessed.

How many of the 16 reprocessed ureteroscopes that researchers tested had dangerously high levels of organic material and bioburden remaining on them? All of them — despite undergoing manual cleaning and sterilization.

That's right, all 16 flexible ureteroscopes used for kidney stone removal and other procedures at 2 facilities had visible irregularities and residual contamination exceeding benchmarks for clean endoscopes, according to alarming study results published in the American Journal of Infection Control and presented at this week's meeting of the Association for Professionals in Infection Control and Epidemiology.

Two-thirds of the scopes contained hemoglobin and nearly half had high levels of adenosine triphosphate (ATP). Visual inspection of the scopes turned up debris in channels, oily deposits and a mystery white, foamy residue the researchers had never before seen.

This study underscores the importance of actively monitoring reprocessing outcomes to ensure reprocessed flexible ureteroscopes are sterile and safe for patient use, says lead study author Cori Ofstead, MSPH, of Ofstead & Associates, a healthcare consulting firm in Saint Paul, Minn.

"If [the reprocessing staffs] had been conducting cleaning verification tests or visual inspections, every one of these scopes would have been removed from service," says Ms. Ofstead. "Every scope should be inspected, every time — and there must be accountability measures set in place."

There should have been no trace of visual bacteria and microbial cultures should have been negative after the scopes were sterilized, note the researchers. They say the scopes involved in the study did not undergo bedside cleaning in the OR and sat in the reprocessing area for long periods before they were cleaned and sterilized with hydrogen peroxide gas, factors that could have contributed to the worrisome findings.

"Sterilization doesn't work if scopes aren't clean," says Ms. Ofstead.

Daniel Cook


Also in the News...

Design Flaw Could Keep Bair Hugger Warming Blankets From Fully Inflating
Pentax Voluntarily Recalls ED-3490TK Video Duodenoscopes for Design and Labeling Changes
Police: N.Y. Surgeon Choked Nurse With Drawstring From His Sweatshirt
Joint Commission Has Zero Tolerance for Poor Hand Hygiene
Kentucky Plastic Surgeon Arrested for Allegedly Arriving at Hospital Intoxicated
Lawsuit: VA Hospital Left Scalpel Inside Army Vet's Abdomen
Single-OR Facilities Must Adhere to the Same Quality Standards as Licensed ASCs

New to Outpatient Surgery Magazine?
Sign-up to continue reading this article.
Register Now
Have an account? Please log in:
Email Address:
  Remember my login on this computer

advertiser banner

Other Articles That May Interest You

Lessons in Lumened Instrument Cleaning

Follow these tips to ensure reprocessing's critical first step is performed correctly each and every time.

5 Strategies for Whole-Room Disinfection

When time and resources are limited, efficiency is the key.

Infection Prevention: Help Your Reprocessors Do the Right Thing

Give your techs the training and the tools they need to bust those bugs.