A Look Inside Low-Temp Sterilization

6 points to keep in mind about reprocessing heat-sensitive instruments.

Kevin Anderson

Kevin Anderson, BSN, RN, CSSM, CRCST, CER

BIO

Nasal Injection
UNDER AN HOUR Shorter cycle times is one key benefit vaporized hydrogen peroxide sterilization has over over ethylene oxide.

Besides not knowing how to get there, the biggest mystery about the sterile processing department might be low-temperature sterilization. Allow me to clear the air about how heat- and moisture-sensitive instruments get back into your ORs, sterile and ready to use on the next patient — but perhaps not as quickly as you'd like.

1. Why low-temp? Nothing destroys bacteria, fungi and spores like steam sterilization, which is also the most cost-effective and simplest sterilization method. But many of today's heat-sensitive instruments can't withstand the higher temperatures or moisture of steam. Enter low-temp sterilization.

2. EtO vs. VHP. Your main choices in low-temp sterilization processes are ethylene oxide (EtO) or vaporized hydrogen peroxide (VHP). Besides being toxic, carcinogenic and explosive, EtO gas has an excessively long sterilization cycle that's incompatible with the pace of outpatient surgery. Including the required aeration to remove EtO residue, the EtO cycle ranges from 14 to 30 hours. Very impractical. Who's got that kind of time to reprocess instruments when the most efficient and safest form of low-temp sterilization that eliminates contaminating microbes in a fraction of the time is available?

With VHP, the most common cycle is 28 minutes — with no aeration necessary. Some items such as lumened flexible scopes may stay in for an hour-long extended cycle. It's simple to operate, install (all you need is a 208 V outlet) and monitor. It's also safe for the environment and compatible with most medical devices. Here's how it works: The sterilization chamber is evacuated and hydrogen peroxide solution is injected from a cassette and vaporized in the sterilization chamber.

3. EtO hasn't gone away. I'm the manager of the sterile processing department at Henry Ford Wyandotte (Mich.) Hospital. We service 11 ORs, 3 procedure rooms and 18 off-site clinics. We reprocess 2,500 to 3,000 instrument trays per month. In addition to 3 steam sterilizers, we have 2 VHP sterilizers and 2 EtO sterilizers. Surprised? Let me explain.

Some instruments are only EtO-compatible. If you have an outdated endoscope inventory, for example, certain flexible scopes might not be validated with VHP. If given the choice to upgrade their instrument inventory or keep their EtO sterilizer in good working order, many OR managers would choose to keep the old instruments in play. With flexible scopes costing the same as a car ($30,000) and low-temp sterilizers ranging in price from $80,000 for certified pre-owned to $180,000 for new, it's something to consider. Side note: EtO could extend the life of your heat- or moisture-sensitive instruments because it's non-corrosive to plastic, metal and rubber materials.

New to Outpatient Surgery Magazine?
Sign-up to continue reading this article.
Register Now
Have an account? Please log in:
Email Address:
  Remember my login on this computer

advertiser banner

Other Articles That May Interest You

Scratch Beneath the Surface

Audit the effectiveness of OR cleaning by taking a big-picture view of the entire process.

New Formula Can Predict Efficacy of Antibiotic Combinations

Souped-up Surface Cleaning

Whole-room disinfection solutions and fine-tuned manual scrubbing will eradicate infection-causing microorganisms.